5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

764

■■■■■■■■

tiamo 2.1 (ProcessLab)

Titrando
Range

1 ... 999 µg/min

Default value

20 µg/min

Drift correction

If this parameter is set to automatic or manual the drift correction ( = Drift
value
× Time for drift correction) caused by empty consumption during
titration is deducted from the total consumption. The Time for drift cor-
rection
is the period between end of conditioning and end of titration.

Selection

automatic | manual | off

Default value

off

automatic
The drift lastly determined during conditioning will be used as drift value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.

Drift value

Drift value for drift correction. This parameter is only shown for Drift cor-
rection
= manual.

Range

0.0 ... 99.9 µL/min

Default value

0.0 µL/min

Stop volume

Determination stops when the specified volume has been dosed since the
start of the conditioning. The stop volume should be adjusted to the size of
the titration cell in order to prevent any overflow.

Titrando
Range

0.000000 ... 9999.99 mL

Default value

20.00 mL

Selection

off

Default value

off

Stop time

Determination stops when the specified time has elapsed since the start of
the conditioning.

Titrando
Range

1 ... 999999 s

Selection

off

Default value

off