5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
464
■■■■■■■■
tiamo 2.1 (ProcessLab)
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 855
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched off automatically when
the command has finished. This parameter is displayed only for Titrandos
and 855.
Switch on/off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched on automatically at the
start of the command and will be switched off automatically at the end of
the command. This parameter is displayed only for 751, 785, 798 and 799.
5.6.3.2.3.3
DET U - Start conditions
Tab: Method
▶ DET U ▶ Properties... ▶ Start conditions
Command name
Name of the command.
Input
25 characters
The start conditions are processed in the listed sequence before the titration
is started.
Initial measured value
Note
Is displayed only for Titrando and 855.
The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters:
Measured value drift
The measured value is only accepted if the drift is less than the value entered
here.
Range
0.1 ... 999.0 mV/min