6 Configuration

tiamo 2.1 (ProcessLab)

■■■■■■■■

1295

Warning limits for slope (only for pH and ISE electrodes)
on | off (Default value: off)

If this option is switched on, then the values for the slope will be shown
in orange if the limits are exceeded.

Lower limit

Lower warning limit for the slope.

for pH electrode
Range

-2.147E+9 ... 2.147E+9 %

Default value

97.5 %

for ISE (+1)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

57.2 mV

for ISE (-1)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-61.2 mV

for ISE (+2)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

27.7 mV

for ISE (-2)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-31.7 mV

for ISE (+3)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

17.7 mV

for ISE (-3)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-21.7 mV

for ISE (+4)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

12.8 mV

for ISE (-4)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-16.8 mV