beautypg.com

Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 1146

background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1130

■■■■■■■■

tiamo 2.1 (ProcessLab)

Set lines

Output signal

Input of the binary pattern for the output signal or selecting a predefined
signal pattern.

It is possible to enter the following characters:

0 = line inactive

1 = line active

* = any line status

p = set pulse (pulse length = 200 ms). If an impulse with a different length
is to be emitted, a corresponding template must be defined.

Titrando, 855, 856, 867, 751, 756, 758, 784, 785, 795, 798, 799, 831,
Sample Processor
Selection

Bit-pattern with exactly 14 characters (0, 1, *,
p)
| ************** | Signal pattern

Default value

**************

702, 716, 718, 719, 720, 721, 736, 794
Selection

Bit-pattern with exactly 8 characters (0, 1, *,
p)
| ******** | Signal pattern

Default value

********

Spectrometer
Selection

Bit-pattern with exactly 10 characters (0, 1, *,
p)
| ********** | Signal pattern

Default value

**********

The bits are numbered from right to left:

Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0

Bit

13 12 11 10 9 8 7 6 5 4 3 2 1 0

Example:

************1* sets the output line 1 to enabled (= set), which would result
in a stop command for a connected Titrino, for example. ************0*
sets the line to inactive.

Note

We recommend masking the irrelevant output lines with an asterisk * so
as not to modify these line conditions.