beautypg.com

Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 837

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

821

Lower hysteresis pH

Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value, then the event Measured value lower
limit ok
is triggered.

Titrando, 855
Range

0.000 ... 20,000

Default value

0.020

Upper limit pH

Upper limit of the measured value. If the measured value exceeds this limit,
then the event Measured value upper limit exceeded is triggered.

Titrando, 855
Range

-20,000 ... 20,000

Default value

20,000

Titrino
Range

-20.00 ... 20.00

Default value

20.00

Upper hysteresis pH

The upper hysteresis of the measured value. If the measured value once
again falls below the upper limit by this hysteresis value, then the event
Measured value upper limit ok will be triggered.

Titrando, 855
Range

0.000 ... 20,000

Default value

0.020

Action

Selection of the event to be triggered if the upper or lower limit is exceeded:

Selection

Stop determination | Stop command (only
Titrando)
| Wait for [Continue] | Wait for limit
ok
| none

Default value

none

Stop determination
The running STAT command is quit, then the exit track (if present) is
started and the determination is stopped.
Stop command (only Titrando)
The running STAT command is quit, then the next command is executed.