beautypg.com

Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 961

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

945

off
No stop.

Temperature

Measurement temperature

Temperature of the solution to be measured. If a temperature sensor is con-
nected, the temperature entered here is ignored.

712
Range

-170.0 ... 500.0°C

Default value

20.0

856
Range

-20.0 ... 150.0°C

Default value

20.0

Reference temperature

The electrical conductivity depends greatly on the temperature. The con-
ductivity measured at a particular temperature is thus automatically conver-
ted to this reference temperature (usually 20 °C or 25 °C).

712
Range

-170.0 ... 500.0°C

Default value

20.0

856
Range

-20.0 ... 150.0°C

Default value

20.0

Selection

off

off
The conductivity is displayed for the measurement temperature.

Temperature compensation
Selection

Temperature coefficient | Sample solution

Default value

Temperature coefficient

Temperature coefficient
Manual input of a constant temperature coefficient.
Sample solution
Selection of a function for the temperature coefficients from the table
Sample solutions (TC conductivity).

Temperature coefficient

Range

0.00 ... 9.99 %/°C

Default value

2.00 %/°C