3 templates for output lines, 1 managing templates for output lines, 2 editing templates for output lines – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1241: Output lines
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.1 (ProcessLab)
■■■■■■■■
1225
1 = Line active and
* = Any line status.
Selection
Bit-pattern with 8 characters (0, 1, *) | ********
Default value
********
The input lines and bits are numbered from right to left:
Input 7 6 5 4 3 2 1 0
Bit
7 6 5 4 3 2 1 0
Example: *******1 expects an active input line 0 (1 = set). For example, this
line is set by a Titrino after a titration has been completed and the Titrino
can accept a start signal again.
Note
Input lines that are of no interest or for which no defined condition can
be predicted should also be masked with an asterisk *.
6.3.3.3
Templates for output lines
6.3.3.3.1
Managing templates for output lines
Dialog window: Configuration
▶ Tools ▶ Templates ▶ Output lines... ▶ Templates
for output lines
The client-specific bit-pattern for setting remote output signals can be
defined in the dialog window Templates for output lines; these can be
selected with the command CTRL and in Manual Control. The table with
the defined templates cannot be edited, although it can be sorted according
to the selected column in increasing or decreasing sequence by clicking on
the column title
[New]
Generates a new template (see Chapter 6.3.3.3.2, page 1225).
[Edit]
Edits the selected template (see Chapter 6.3.3.3.2, page 1225).
[Delete]
Deletes the selected template.
6.3.3.3.2
Editing templates for output lines
Dialog window: Configuration
▶ Tools ▶ Templates ▶ Output lines... ▶ Templates
for output lines
▶ [New] / [Edit] ▶ New template / Edit template
With [New] or [Edit] the dialog window New template or Edit tem-
plate opens for entering a new template or for editing an existing template.