5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1012
■■■■■■■■
tiamo 2.1 (ProcessLab)
Parameters
The parameters for the command CAL MEAS pH are configured on the
following 2 tabs:
■
General/Hardware
Parameters for devices, sensors and stirrers.
■
Measuring parameters
Parameters for setting the measurement procedure.
Command variables
The following command variables are generated in the method run by the
command CAL MEAS pH and can be used in formulas under the designa-
tion 'Command name.Variable designation':
Identifica-
tion
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
not ever been started
.DBL
Total duration for the processing of the command in s
.EME
Final measured value (measured value after processing of
the command) in the unit of the measured value
.ETE
End temperature (temperature following processing of
the command) in °C
.FIN
Command status; 1 = Command has been ended at
least once, 0, invalid (variable not available) = Com-
mand has never been ended
.IME
Initial measured value (measured value before start con-
ditions are processed) in the unit of the measured value
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.LP.MEA
Measured value for the last measuring point in the unit
of the measured value
.LP.TEM
Temperature for the last measuring point of the measur-
ing point list in °C
.LP.TIM
Time in s until the last measuring point in the measuring
point list is reached