6 Configuration

tiamo 2.1 (ProcessLab)

■■■■■■■■

1287

Shared calibration data

Calibration temperature

Temperature during the calibration. The display also shows whether the
temperature was measured with a Pt1000 or an NTC temperature sensor or
entered manually.

Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Calibration date

Date and time of last calibration that is entered automatically after each
automatic calibration or manual entry.

Calibration method

Name of the method with which the last calibration was carried out. If the
calibration data was entered manually then manual will appear here.

User

Short name of the user logged in during the calibration or when the cali-
bration data was entered manually. If work is not carried out using log in
then the user logged in under Windows will be entered automatically.

Measuring input

Display of the type and the serial number of the measuring input at which
the sensor was connected during the calibration (only for Titrando).

[Initial data]

Opens the dialog window Initial data sensor (see Chapter 6.7.3.6, page
1294)
, in which the initial calibration data stored in the sensor chip is dis-
played (only for intelligent sensors).

Calibration data monitoring
on | off (Default value: off)

If this option is switched on then the validity of the calibration will be moni-
tored.

Calibration interval

Validity of the calibration in days. If a value is entered here then the date for
the Next calibration will be adapted automatically.