5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1062
■■■■■■■■
tiamo 2.1 (ProcessLab)
[Delete]
Deletes the calculated measured value selected in the table.
Additional external measured values
Additional external measured values
on | off (Default value: off)
If this option is enabled, then a maximum of 3 new measured values can be
taken over and saved from existing measuring commands (i.e. MEAS) which
can also be displayed in curves under the designation Extern 1…3 and used
in formulas as variables 'Command name.EX1…3'-
[New]
Opens the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.10.2, page 859).
[Properties]
Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.10.2, page 859).
[Delete]
Deletes the external measured value selected in the table.
5.6.6.3.5
DOS U
5.6.6.3.5.1
DOS U - Overview
Dialog window: Method
▶ DOS U ▶ Properties... ▶ DOS U - 'Command name'
Command for controlled dosing with measured value U.
Devices
This command can be executed with the following devices:
Titrando: 835, 836, 842, 857, 902, 906, 907
Titrino: 718, 736, 751, 799
pH/Ion Meter: 867
Robotic Titrosampler: 855
Appearance
The command has the following appearance: