3 command variables, Command variables – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 44
2.4 Formula editor
■■■■■■■■■■■■■■■■■■■■■■
28
■■■■■■■■
tiamo 2.1 (ProcessLab)
Variable name
Description
Data type
.OVF
Exceeding limits for method
variable (Number: 1 = limit
exceeded, 0 = limit not
exceeded)
Number
In the following lines you will find the method variables (sample data)
present in the default settings which appear in the workplace and which
can be edited and deleted in the START command of the corresponding
method.
Sample size.VAL
Value of 'Sample size'
Number
Sample size.OVF
Exceeding limit for 'sample
size'
Number
Unit.VAL
Value of 'Unit'.
Text
Unit.OVF
Exceeding limit for 'unit'
Number
Sample posi-
tion.VAL
Value of 'Sample position'
Number
Sample posi-
tion.OVF
Exceeding limit for 'sample
position'
Number
ID1 (…3).VAL
Value of 'ID1 ... 3'
Text
ID1 (…3).OVF
Exceeding limit for 'ID…3'
Number
2.4.3.3
Command variables
Dialog window: Formula Editor
▶ Variables
The command variables are method-specific, that depend on which com-
mands are used in the method. Included among the command variables are
also solution and sensor variables, which are applied automatically from the
corresponding tables in the Configuration for the device-dependent com-
mands at the time the determination is started and which are allocated to
the individual commands. The Variables field of the Formula editor lists all
Command variables which are available for the current method.
Syntax
'Command name.Variable identification'
Examples: 'DET U 3.SME', 'Track 6.BSY', 'Liquid Handling 4.CONC'
You can select the command variables directly in the formula editor under
Variables
▶ Command Variables in order to avoid syntax errors.
Command variables
Unless marked otherwise, all of the variables listed here in alphabetical order
are of the Number type.