5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

1081

Exchange position
With this function the cylinder is filled via the given port first. The valve
disc is then turned to Port 2 and the dosing unit can be removed from
the dosing drive.
Change port
With Change port the valve disc is only moved to the given port; no
piston movement takes place.
Compensate
As the dosing units are exchangeable, the coupling of the Dosino piston
rod (spindle) has a low mechanical tolerance that can be noticed when
the piston changes its direction of movement. This tolerance can be
compensated for with the Compensate function. A short piston move-
ment is first made in the same direction as the previous movement, which
is then followed by a piston movement in the reverse direction.

Port

The port where each function is to be carried out needs to be defined for
each liquid handling command.

Selection

1 | 2 | 3 | 4

1
Default setting for Dosing and Aspirate
2
Default setting for Fill, Exchange position and Change port
4
Default setting for Eject to stop, Eject to end volume and Compensate

Volume

Enter the volume to be moved. This parameter is available only for the
Functions Dosing or Aspirate.