5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

1087

Dosing device

Dosing device

Selection of the dosing device (Dosinos only) to be emptied. All the dosing
device connections which are possible with the selected device type are dis-
played.

Titrando, 846, 856, 867
Selection

1 | 2 | 3 | 4

Default value

1

814, 815, 855, 864, 874
Selection

1 | 2 | 3

Default value

1

730, 774
Selection

1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12

Default value

1

Titrino
Selection

external D1 | external D2

Default value

external D1

Solution

Input of a solution name or selection of a solution from the solutions defined
in the solution table. If intelligent exchange or dosing units are used, then
a check will be made in the method sequence to see whether the correct
solution has been set on the selected dosing device and whether the dosing
device type is correct. With non-intelligent exchange or dosing units, only
the cylinder volume is checked. At the start of the command, a check is made
of the working life, the validity of the titer and the GLP test interval for the
selected solution.

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.