Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 893
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
877
Selection
Sensor name | pH electrode | Metal electrode |
ISE electrode
Default value
Metal electrode
Temperature measurement
Type of temperature measurement.
Titrando, 855, 867
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise, the temperature entered manually under
Temperature on the Measuring parameters tab will be used.
off
The temperature will not be measured. The temperature manually
entered under Temperature on the Measuring parameters tab will
be used.
Stirrer
Stirrer
Selection of the stirrer.
Titrando, 855, 867
Selection
1 | 2 | 3 | 4 | off
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 855, 867
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched off automatically when
the command has finished. This parameter is displayed only for Titrandos,
855 and 867.