5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
844
■■■■■■■■
tiamo 2.1 (ProcessLab)
5.6.3.8.7.6
STAT U - Stop conditions
Tab: Method
▶ STAT U ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Input
25 characters
Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop the
titration.
Stop volume
Stops when the given volume has been added after the start of the titration
(including start conditions). The stop volume should be adapted to suit the
sample weight or the titration vessel size.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
100,000 mL
Selection
off
Titrino
Range
0.000 ... 9999.99 mL
Default value
100.00 mL
Selection
off
Stop time
Stops when the preset time has elapsed after the start conditions have been
fulfilled.
Range
0 ... 999999 s
Selection
off
Default value
off
since
Specifies when the stop time begins to be counted down (will not be dis-
played if the Stop time is disabled):
Selection
Start | Control point first reached | Last dosing
Default value
Start
Start
The stop time starts after the start volume has been dosed and after the
following pause time.
Control point first reached
The stop time starts as soon as the control point has been reached for
the first time.