3.6 Subwindow Run
■■■■■■■■■■■■■■■■■■■■■■
122
■■■■■■■■
tiamo 2.1 (ProcessLab)
Sample position
Position of the sample on the rack. This number is used to move to the
sample position with the command MOVE. For this to be achieved, the
Sample position option must be set under Target for the Move para-
meter.
Range
1 ... 999 (Increment: 1)
Default value
'empty'
ID1…ID16
Sample identifications ID1…ID16. Any text may be entered into this field,
the checking for type and limit values does not take place until the start of
the determination. For sample identifications of the type Date/Time, the
entry must be made in the format YYYY-MM-DD or YYYY-MM-DD
hh:mm:ss. If text templates are defined for the sample identification, then
these can be selected (see Chapter 3.5.1, page 115).
Input
100 characters
Selection
Selection of text templates
Sample size
Sample size (weigh-in). Only numbers, decimal points, +, –, E and e can be
entered in this field. The checking for limit values does not take place until
the start of determination.
Note
If a value is entered or imported into this field, then in addition the time
of the data entry and the data source are stored manually, while the
'Name of the balance', the 'Name of the barcode reader' or the
'Name of the import file) are stored automatically at the same time,
placed in the determination and identified in the Information sub-
window as Entry date and Data source.
Note
If a negative value is entered in this field (e. g. during data import from a
balance in the event of reweighing), then this will be also be calculated
as a negative in the Formula Editor.
Input
Number with a maximum of 10 digits
Default value
1.0