3 digital in, 1 digital in - overview, Digital in – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1184
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1168
■■■■■■■■
tiamo 2.1 (ProcessLab)
■
>0: For 100% stirrer output (Parameter 1 = 100), pulse duration in 10
ms
Example:
Parameter 1 = 100, Parameter 2 = 30: Pulse duration switched-on
300 ms, switched off 0 ms
Parameter 1 = 75, Parameter 2 = 30: Pulse duration switched-on 225
ms, switched off 75 ms
Range
-214783647 ... 214783647
Default value
0
5.6.11.3
DIGITAL IN
5.6.11.3.1
DIGITAL IN - Overview
Dialog window: Method
▶ DIGITAL IN ▶ Properties... ▶ DIGITAL IN - 'Command
name'
Devices
This command can be executed with the IO Controller and with the Stepping
Motor Controller.
Appearance
The command has the following appearance:
Mode of operation
The statuses of all digital inputs will be entered and saved in command var-
iables at the time of the command (syntax Command name.Variable
designation)
Parameters
The parameters for the command DIGITAL IN are configured on the fol-
lowing 2 tabs:
■
General
Parameters for device.
■
Stop criteria
Parameters for defining the stop criteria.
Command variables
The following command variables are generated in the method run by the
command DIGITAL IN and can be used in formulas under the designation
'Command name.Variable designation':