6.7 Subwindow Sensors

■■■■■■■■■■■■■■■■■■■■■■

1296

■■■■■■■■

tiamo 2.1 (ProcessLab)

Upper limit

Upper warning limit for the slope.

for pH electrode
Range

-2.147E+9 ... 2.147E+9 %

Default value

102.5 %

for ISE (+1)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

61.2 mV

for ISE (-1)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-57.2 mV

for ISE (+2)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

31.7 mV

for ISE (-2)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-27.7 mV

for ISE (+3)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

21.7 mV

for ISE (-3)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-17.7 mV

for ISE (+4)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

16.8 mV

for ISE (-4)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-12.8 mV

Intervention limits for slope (only for pH and ISE electrodes)
on | off (Default value: off)

If this option is switched on, then the values for the slope will be shown
in red if the limits are exceeded.