5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
657
Titrino
Range
-170.0 ... 500.0 °C
Default value
25.0 °C
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
Titrando, 855
Range
0.1 ... 999999 s
Default value
2.0 s
Titrino
Range
0.08 ... 16200 s
Default value
2.00 s
5.6.3.4.3.6
SET pH - Stop conditions
Tab: Method
▶ SET pH ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Input
25 characters
The conditions for stopping the titration are defined on this tab. If more than
one stop condition is in effect at the same time, then the criterion which is
fulfilled first will stop the titration.
Stop volume
Stops when the given volume has been added after the start of the titration
(including start conditions). The stop volume should be adapted to suit the
sample weight or the titration vessel size.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
100,000 mL
Selection
off
Titrino
Range
0.000 ... 9999.99 mL
Default value
100.00 mL
Selection
off
Stop time
Stops when the preset time (including start conditions) has elapsed since the
start of the titration.