5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
938
■■■■■■■■
tiamo 2.1 (ProcessLab)
■
Measuring parameters
Parameters for setting the measurement procedure.
■
Evaluations
Defining further methods for evaluating measurement curves.
■
Additional measured values
Defining further measured values which can be saved as additional col-
umns in the measuring point list.
Command variables
The following command variables are generated in the method run by the
command MEAS Cond and can be used in formulas under the designation
'Command name.Variable designation':
Identification
Description
.BP{x}.MEA
Measured value for the break point (1…9) in the unit
of the measured value
.BP{x}.TEM
Temperature for the break point × (1…9) in °C
.BP{x}.TIM
Time for the break point × (1…9) in s
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
.CLC
Cell constant of the sensor used in the command for
conductivity measurement cells
.DBL
Total duration for the processing of the command in s
.EME
End measured value (measured value after processing
of the command) in the unit of the measured value
.ENP
Electrode zero point of the sensor used in the command
.ETE
End temperature (temperature following processing of
the command) in °C
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command
has never ended
.FP{x}.MEA
Measured value for the fixed endpoint × (1…9) in the
unit of the measured value
.FP{x}.TEM
Temperature for the fixed endpoint × (1…9) in °C
.FP{x}.TIM
Time in s until the fixed endpoint × (1…9) is reached
.IME
Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value