16 stdadd auto, 1 stdadd auto - overview, Stdadd auto – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1001
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
985
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Range
0.1 ... 999999.0 s
Default value
2.0 s
Temperature
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous, then
the temperature will be measured continuously.
Range
-20.0 ... 150°C
Default value
25.0
Switch off stirrer during measurement
Switch off stirrer during measurement
on | off (Default value: off)
If this option is activated, then the stirrer is switched off during the meas-
urement. In this case the following two parameters can be edited:
Stir before measurement
Stirring time between end of dosing and measurement.
Range
0 ... 999999 s
Default value
0 s
Pause before measurement
Waiting time before the measurement during which no stirring takes place.
Range
0 ... 999999 s
Default value
0 s
5.6.4.16
STDADD auto
5.6.4.16.1
STDADD auto - Overview
Dialog window: Method
▶ STDADD auto ▶ Properties... ▶ STDADD auto - 'Command
name'
Command for Standard addition with automatic addition of the standard
addition solution from a dosing device in such a way that a constant poten-
tial difference results.
Devices
This command can be executed with the following devices: