Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1307
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.1 (ProcessLab)
■■■■■■■■
1291
for ISE (-2)
Range
-2.147E+9 ... 2.147E+9 mV
Default value
-25.0 mV
for ISE (+3)
Range
-2.147E+9 ... 2.147E+9 mV
Default value
23.0 mV
for ISE (-3)
Range
-2.147E+9 ... 2.147E+9 mV
Default value
-16.0 mV
for ISE (+4)
Range
-2.147E+9 ... 2.147E+9 mV
Default value
17.0 mV
for ISE (-4)
Range
-2.147E+9 ... 2.147E+9 mV
Default value
-12.0 mV
pH(0) monitoring (for pH electrodes only)
on | off (Default value: off)
If this option is switched on then a check will be made to see whether the
electrode zero point pH(0) determined during the calibration of a pH elec-
trode is within the defined limits. If it is outside them, a corresponding
message appears and you can choose whether you will nevertheless accept
the value or whether the previous value is to be retained.
Lower limit
Lower limit for pH(0).
Range
-2.147E+9 ... 2.147E+9
Default value
6.750
Upper limit
Upper limit for pH(0).
Range
-2.147E+9 ... 2.147E+9
Default value
7.250
E(0) monitoring (for ISE electrodes only)
on | off (Default value: off)
If this option is switched on then a check will be made to see whether the
electrode zero point E(0) determined during the calibration of an ISE elec-
trode is within the defined limits. If it is outside them, a corresponding