6.2 Administration
■■■■■■■■■■■■■■■■■■■■■■
1202
■■■■■■■■
tiamo 2.1 (ProcessLab)
Security settings can be exported and imported. This means that these set-
tings can be exchanged between different client/server systems.
The security settings can be put out as PDF file using the menu item
File
▶ Print (PDF) ▶ Security settings.
Tabs
The security settings are configured on the following tabs:
■
Login/Password protection
Settings for login and password protection.
■
Audit Trail/Changes
Enable/Disable the Audit Trail and comments when methods, determi-
nations or sample data are modified.
■
Signatures
Options for electronic signatures.
■
Default reasons
Definition of reasons for signing and modifying methods, determinations
and sample data.
6.2.2.2
Login/Password protection
Tab: Configuration
▶ Tools ▶ Security settings ▶ Login/Password protection
On the tab Login/Password protection is defined whether the user must
log in with his/her name or with name and password, and how the password
is to be monitored and constructed.
Settings according to 21 CFR 11
[Set]
Sets the parameters to default settings for FDA-compliant work.
Login policies
Enforce login with user name
If this check box is enabled then every time that the program starts, the
login window will appear in which the user must enter his/her user name.
If this check box is disabled no login is necessary and the user logged in
under Windows will be adopted as user name. In this case all subsequent
parameters are deactivated.
Selection
on | off
Default value
off
FDA default: on
Enforce login with password
If this check box is enabled then every time that the program starts, the
login window will appear in which the user has to enter a password as well