9 det ipol - measured value window – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 523
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
507
5.6.3.2.4.9
DET Ipol - Measured value window
Dialog window: Method
▶ DET Ipol ▶ Properties... ▶ Potentiometric evalua-
tion
▶ Evaluation with measured value window (Ipol) ▶ [New]/[Proper-
ties]
▶ Measured value window
Measured value windows are regions on the measured value axis for
which different parameters for potentiometric evaluation can be defined.
Only those endpoints which lie within these windows and additionally meet
the parameters defined for each window will be recognized. Only one end-
point will be recognized per window.
Lower limit
Lower limit for the measured value window.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Default value
-2000.0 mV
Titrino
Range
-2000 ... 2000 mV
Default value
-2000 mV
Upper limit
Upper limit for the measured value window.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Default value
2000.0 mV
Titrino
Range
-2000 ... 2000 mV
Default value
2000 mV
EP criterion
Criterion for the recognition of endpoints. Endpoints whose discovered ERC
value is smaller than the value entered here will not be recognized.
Titrando, 855
Range
0 ... 200
Default value
5
EP recognition
Filter for the recognition of equivalence points: