5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

1119

into the text field (see Chapter 2.5.2, page 79). The formula editor can also
be called up inside the text field (see Chapter 2.4, page 23).

Input

Text (unlimited)

Display message

on | off (Default value: on)

If this option is enabled, all active tracks are stopped and the message
defined in the text box will be displayed in the event that the limits are
exceeded. You can choose whether you want to continue the stopped tracks
with [Continue] or - if defined - to execute the Action listed below.

Record message

on | off (Default value: on)

If this option is enabled, the message defined in the text box will be docu-
mented in the determination if the limits are exceeded. The run will not be
stopped.

Message by E-mail

on | off (Default value: off)

If this option is enabled, the message defined in the text box will be sent to
the address defined under [E-mail] in the event that the result limits are
exceeded.

[E-mail]

Opens the dialog window Send E-mail for defining the E-mail parameters
(see Chapter 2.6.1, page 81).

Acoustic signal

on | off (Default value: off)

If this option is enabled, an acoustic signal will be emitted in addition to the
message defined above when a limit is exceeded.

Action

Action

on | off (Default value: off)

If this option is enabled, one of the following actions will be executed in the
event that limit values are exceeded.

Selection

Stop determination | Stop determination and
series

Default value

Stop determination