Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 746
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
730
■■■■■■■■
tiamo 2.1 (ProcessLab)
Identification
Description
.EVT
End volume (total dosed volume at the end of the com-
mand) in mL
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command
has never ended
.FP{x}.MEA
Measured value for the fixed endpoint × (1…9) in mV
.FP{x}.TEM
Temperature for the fixed endpoint × (1…9) in °C
.FP{x}.TIM
Time in s until the fixed endpoint × (1…9) is reached
.FP{x}.VOL
Volume for the fixed endpoint × (1…9) in mL
.IME
Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.LP.CAx
Calculated value × (1...3) for the last measuring point
on the measuring point list
.LP.DVT
dV/dt for the last measuring point in the measuring
point list
.LP.EXx
External value × (1...3) for the last measuring point of
the measuring point list
.LP.MEA
Measured value for the last measuring point in the
measuring point list in the unit of the measured value
.LP.TEM
Temperature for the last measuring point of the meas-
uring point list in °C
.LP.TIM
Time in s until the last measuring point in the measuring
point list is reached
.LP.VOL
Volume for the last measuring point in the measuring
point list in mL
.MA.MEA
Maximum measured value in the unit of the measured
value
.MA.TEM
Temperature for the maximum measured value in °C
.MA.TIM
Time in s until the maximum measured value is reached
.MA.VOL
Volume with maximum measured value in mL
.MI.MEA
Minimum measured value in the unit of the measured
value