6 Configuration
tiamo 2.1 (ProcessLab)
■■■■■■■■
1259
6.6.3.2
Properties - Solution
Tab: Configuration
▶ Titrants/Solutions ▶ [Edit] ▶ Proper-
ties...
▶ Solution - 'Name' ▶ Solution
Solution name
Name of the solution (can be entered or selected).
Input
24 characters
Selection
Selection from predefined names
Concentration (value)
Concentration value of the solution. This value for the concentration of a
solution to be used in the titration commands DET, MET, SET and KFT, is
available as a variable ''Command name.CONC' for calculations.
Range
1E-12 ... 1E+12 (max. 10 digits)
Default value
1.000
Concentration (unit)
Concentration unit of solution.
Input
10 characters
Selection
mol/L | mmol/L | µmol/L | g/L | mg/L | µg/L |
mg/mL | ppm | % | mEq/L
Default value
mol/L
Comment
Remarks about the solution (e.g. batch number.)
Input
24 characters
Production date
Date on which the solution was produced. This date can be edited only for
manually added solutions by pressing
.
Selection
Date selection
Solution monitoring
on | off (Default value: off)
If this option is switched on then the working life of the solution will be
monitored.
Working life
Working life of the solution in days. If a value is entered here then the Expiry
date will be automatically adapted.