1.10 What is new in tiamo 2.1?

■■■■■■■■■■■■■■■■■■■■■■

10

■■■■■■■■

tiamo 2.1 (ProcessLab)

Titrino: The window evaluation does not work correctly.

778, 789: Exception if the lift is to be lowered at an invalid rack position.

Sartorius balances: Transmit data only when using point as decimal sep-
arator.

Commands

DET: Difficulties to correctly transmit the max. increment to the device
when changing the type of device (Titrino/Titrando) during the com-
mand.

CTRL/SCAN with devices with less than 14 output lines (Titrino,
Avantes): The lines were counted from the left instead of from the right

MEAS Cond: Exception, if Reference temperature = off is being set.

MEAS Cond: Additional calculated values are not shown in the live curve
window when the result unit was selected from the list.

KFT, SET, KFC: The start drift is 0.0 when the automatic conditioning
is disabled.

CAL LOOP pH: Calibration using special buffer doesn’t work when using
manual temperature entry different from 25°C.

Database program part

Unexpectedly steep EPs (ERC > 1000) are being lost during recalculation.

Method program part

Exception when recalculating a calculation with CV = invalid.

Configuration program part

Exception in the Audit Trail, if "date#" is being used as an entry (e.g. ID1).