8.4 Manual Control - Functions
■■■■■■■■■■■■■■■■■■■■■■
1488
■■■■■■■■
tiamo 2.1 (ProcessLab)
Titrino
Range
-127 ... 127 µA (Increment: 1)
Default value
5 µA
U(pol)
The polarization voltage is the potential applied to the polarized electrode
during an amperometric measurement.
Titrando, 855
Range
-1250 ... 1250 mV (Increment: 25)
Default value
400 mV
Titrino
Range
-1270 ... 1270 mV (Increment: 10)
Default value
400 mV
Reference temperature
The electrical conductivity depends greatly on the temperature. The con-
ductivity measured at a particular temperature is thus automatically conver-
ted to this reference temperature (usually 20 °C or 25 °C).
712
Range
-170.0 ... 500.0 °C
Default value
20.0
856
Range
-20.0 ... 150.0 °C
Default value
20.0
Selection
off
off
The conductivity at the measuring temperature is displayed.
Temperature compensation
Selection
Temperature coefficient | Sample solution
Default value
Temperature coefficient
Temperature coefficient
Manually enter a constant temperature coefficient.
Sample solution
Select a function for the temperature coefficient from the table Sample
solutions (TC conductivity).
Temperature coefficient
Range
0.00 ... 9.99 %/°C