Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 786
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
770
■■■■■■■■
tiamo 2.1 (ProcessLab)
■
Additional evaluations
Definition of additional methods for the evaluation of the titration curves.
■
Additional measured values
Definition of additional measured values of other measuring commands
which are saved as additional columns in the measuring point list.
Command variables
The following command variables are generated in the method run by the
command KFC and can be used in formulas under the designation 'Com-
mand name.Variable designation':
Identification
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
.CBY
Command status; 1 = Conditioning activated, 0 =
Conditioning deactivated
.COK
Command status; 1 = Condition requirement ful-
filled, 0 = Condition requirement not fulfilled
.DBL
Total duration for the processing of the command in s
.DRI
Current and/or last drift for drift correction in µL/min
.DTI
Time for drift correction (time from the start of the
titration to the end of the command) in s
.EME
End measured value (measured value after processing
of the command) in the unit of the measured value
.EP.CHA
Load for endpoint in mA·s
.EP.DVT
Drift for the endpoint in µg/min
.EP.MEA
Measured value for the endpoint in mV
.EP.QTY
Measured value (water) for the endpoint in µg
.EP.TEM
Temperature for the endpoint 1 in °C
.EP.TIM
Time in s until the endpoint is reached
.EP.TIM
Time in s until the endpoint is reached
.ETE
End temperature (temperature following processing of
the command) in °C
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command
has never ended
.FP{x}.DVT
Drift for the fixed endpoint × (1…9) in µg/min