beautypg.com

Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 908

background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

892

■■■■■■■■

tiamo 2.1 (ProcessLab)

Titrando, 855, 867
Range

0 ... 999999 s

Default value

0 s

Max. waiting time

If the signal drift has not yet been reached, then the measured value will be
accepted when the maximum waiting time has elapsed. If the waiting time
has not been newly entered then a waiting time that is suitable for the drift
will be calculated automatically according to the following equation:

Titrando, 855, 867
Range

0 ... 999999 s

Default value

52 s

Titrino
Range

0 ... 9999 s

Default value

52 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Titrando, 855, 867
Range

0.1 ... 999999.0 s (Increment: 0.1)

Default value

2.0 s

Titrino
Range

0.08 ... 16200 s (Increment: 0.08)

Default value

2.0 s

Stop measured value

Stops when the preset measured value has been reached since the start of
the measurement.

Titrando, 855, 867
Range

-2000.0 ... 2000.0 mV

Selection

off

Default value

off

off
No stop.

Measurement without drift control
If the option Measurement without drift control has been selected, then