Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1419
■■■■■■■■■■■■■■■■■■■■■■
7 Devices
tiamo 2.1 (ProcessLab)
■■■■■■■■
1403
Min. beaker radius
Definition of the minimum radius which the beakers used on the rack must
have. If the lift is moved to the work position the beaker radii defined in
Rack table(see Chapter 6.10.2.1, page 1317) for general Sample posi-
tions(see Chapter 6.10.3.2, page 1320) and Special beakers(see Chapter
6.10.3.4, page 1322) will be compared with the Min. beaker radius. If
this minimum beaker radius is undercut then the run will be stopped and an
error message will appear. With off no check will be made.
Selection
1.0 … 100.0 mm | off
Default value
off
Lift rate
Lift speed for manual operation.
Range
3 ... 25 mm/s
Default value
25 mm/s
Axial distance
Distance between the axis of rotation of the sample rack and swing axis of
the robotic arm.
Range
100.0 ... 300.0 mm
814 USB Sample Processor XL
Selection
166.0 mm
Default value
166.0 mm
815 Robotic USB Sample Processor XL, 855 Robotic Titrosampler, 864 Bal-
ance Sample Processor
Selection
196.0 mm
Default value
196.0 mm
Swing Head
This parameter is only shown when a swing head is mounted on the tower.
Serial number
Shows the serial number of the swing head.
Swing position
Lift position valid for all 4 external positions, at which the swing head turns
to the external positions.
Range
0 ... 235 mm
Default value