5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1034
■■■■■■■■
tiamo 2.1 (ProcessLab)
5.6.5.9.3
CAL Spec - Measuring parameters
Tab: Method
▶ CAL Spec ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The parameters for the measuring procedure can be defined on this tab.
Integration time
Integration time for spectrum recording.
Range
0.01 ... 600000 ms
Default value
6 ms
Averaged spectra
The number of spectra that are recorded and averaged.
Range
1 ... 10000
Default value
10
5.6.5.9.4
CAL Spec - Calibration wavelengths
Tab: Method
▶ CAL Spec ▶ Properties... ▶ Calibration wavelengths
Command name
Name of the command.
Input
25 characters
The reference wavelengths with which the spectrometer is calibrated can
be entered on this tab.
Tolerance
Tolerance for the preset wavelength when assigning the highest peak max-
imum found during the calbration measurement to the reference wave-
length. If a peak is outside the range Wavelength +/- Tolerance, the
calibration will be canceled. The existing calibration in the configuration
remains in effect.
Range
0.1 ... 1100.0 nm
Default value
0.5 nm
[New]
Opens the dialog window Calibration wavelength # to enter wave-
lengths and tolerances (see Chapter 5.6.5.9.5, page 1035).