5 kfc - titration parameters, Titration parameters – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 795
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
779
Range
1 ... 999 µg/min
Default value
5 µg/min
Relative stop drift
The titration is stopped when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift at the titration start and the
relative stop drift defined here. This parameter is only shown for Stop cri-
terion = rel. drift.
Range
0 ... 999 µg/min
Default value
5 µg/min
Delay time
If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is only
shown for Stop criterion = time.
Range
0 ... 999 s
Default value
10 s
5.6.3.6.5
KFC - Titration parameters
Tab: Method
▶ KFC ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Input
25 characters
The parameters for the running of the titration are defined on this tab.
Extraction time
The titration is not stopped until the extraction time has elapsed (even when
the EP has already been reached). The entry of an extraction time may be
advisable, e.g. for the titration of sparingly soluble samples.
Range
0 ... 999999 s
Default value
0 s
Temperature
Manually entered titration temperature.
756, 831
Range
-170.0 ... 500.0 °C
Default value
25.0 °C
851, 852
Range
-20.0 ... 150.0 °C
Default value
25.0 °C