5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

941

Temperature measurement

Type of temperature measurement

856
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise, the temperature entered manually under
Temperature on the Measuring parameters tab will be used.
off
The temperature will not be measured. The temperature manually
entered under Temperature on the Measuring parameters tab will
be used.

Stirrer

Note

Is displayed only if the 856 Conductivity Module has been selected as
device type.

Stirrer

Selection of the stirrer.

Selection

1 | 2 | 3 | 4 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Range

-15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched off automatically when
the command has finished.