5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
831
Identification
Description
.EME
End measured value (measured value after processing
of the command) in the unit of the measured value
.ENP
Electrode slope of the sensor used for the command (in
mV for ISE sensors)
.ETE
End temperature (temperature following processing of
the command) in °C
.EVT
End volume (total dosed volume at the end of the com-
mand) in mL
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command
has never ended
.FP{x}.MEA
Measured value for the fixed endpoint × (1…9) in mV
.FP{x}.TEM
Temperature for the fixed endpoint × (1…9) in °C
.FP{x}.TIM
Time in s until the fixed endpoint × (1…9) is reached
.FP{x}.VOL
Volume for the fixed endpoint × (1…9) in mL
.IME
Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.LP.CAx
Calculated value × (1...3) for the last measuring point
on the measuring point list
.LP.DVT
dV/dt for the last measuring point onthe measuring
point list (SET, KFT, STAT, DOS) or drift for the last
measuring point on the measuring point list in µg/min
(KFC)
.LP.EXx
External value × (1...3) for the last measuring point of
the measuring point list
.LP.MEA
Measured value for the last measuring point in the unit
of the measured value of the measuring point list
.LP.TEM
Temperature for the last measuring point of the meas-
uring point list in °C
.LP.TIM
Time in s until the last measuring point in the measuring
point list is reached
.LP.VOL
Volume for the last measuring point in the measuring
point list in mL