5 calibration commands, 1 calibration commands - overview, Calibration commands – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1009
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
993
Switch off stirrer during measurement
Switch off stirrer during measurement
on | off (Default value: off)
If this option is activated, then the stirrer is switched off during the meas-
urement. In this case the following two parameters can be edited:
Stir before measurement
Stirring time between end of dosing and measurement.
Range
0 ... 999999 s
Default value
0 s
Pause before measurement
Waiting time before the measurement during which no stirring takes place.
Range
0 ... 999999 s
Default value
0 s
5.6.5
Calibration commands
5.6.5.1
Calibration commands - Overview
Menu item: Method
▶ Insert ▶ New command... ▶ Calibrate
Commands for calibration of pH- and ion-selective sensors.
General
The commands CAL LOOP and CAL MEAS are available for calibrating
electrodes. The number and type of buffers or standards, respectively, is
defined in the loop command CAL LOOP, which is to be measured subse-
quently with the command CAL MEAS. After each looping pass, the validity
of the measurement is checked, and at the end the calibration data for the
electrode is calculated.
The basic structure of a calibration method looks as follows: