5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1112

■■■■■■■■

tiamo 2.1 (ProcessLab)

Result parameters
The parameters for the individual results on the results table are configured
on the following 3 tabs:

Definition
Entering the formula, assigning the unit and switching the statistics on/
off.

Monitoring
Settings for result monitoring.

Options
Saving the result as common variable or as titer.

Command variables
The following command variables are generated in the method run by the
command CALC and can be used in formulas under the designation 'Com-
mand name.Variable designation'
:

Identifi-
cation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
not ever been started

.FIN

Command status; 1 = Command has been ended at
least once
, 0, invalid (variable not available) = Com-
mand has never been ended

Result variables
The following result variables are defined in the CALC command and can
be used in formulas under the designation 'RS.ResultName.Variable
designation'
:

Identifica-
tion

Description

.VAL

Result value, optional, i.e. 'RS.RS01' = 'RS.RS01.VAL'

.ASD

Absolute standard deviation for the result

.MAX

Maximum value of the result

.MIN

Minimum value of the result

.MNV

Mean value of the result

.NSR

Statistics actual counter for the result

.NST

Statistics setpoint counter for the result