6.7 Subwindow Sensors

■■■■■■■■■■■■■■■■■■■■■■

1286

■■■■■■■■

tiamo 2.1 (ProcessLab)

Specific calibration data for ISE electrodes

Slope

Slope of the ISE electrode. This value can be determined automatically with
a calibration from the linearized calibration curve or entered manually. The
slope of an ISE electrode is available for calculations as the variable 'Com-
mand name.SLO
.

Range

-999.9 ... 999.9 mV

Default value

59.2/Valency mV

E(0)

E(0) is the second characteristic quantity of the calibration function U =
f(log c)
. E(0) is the y-axis intercept of the calibration curve and corresponds
to the potential at log c = 0. The electrode zero point E(0) of an ISE electrode
is available as a variable 'Command name.ENP' for calculations.

Range

-2000.0 ... 2000.0 mV

Default value

0.0 mV

c (blank)

c (blank) is the third characteristic quantity of the calibration function U =
f(log c)
. It reflects to a certain extent the curvature of the calibration function
at the lowest concentrations. This curvature is caused by the influence of so-
called interfering ions. If the concentration calibration is carried out with less
than 3 standards then c (blank) will be set to zero. The parameter c
(blank)
is available as a variable 'Command name.BLV' for calculations.

Value
Range

0.00 ... 999999999

Default value

0.00

Unit
Input

10 characters

Default value

'empty'

Selection

mol/L | mmol/L | ppm | % | g/L | mg/L | µg/L |
mg/mL | µg/mL

Specific calibration data for conductivity sensors

Cell constant

Cell constant of the conductivity sensor. This value can be determined auto-
matically with a calibration or entered manually. The cell constant is available
as a variable 'Command name.BLV' for calculations.

Range

0.001 ... 500 /cm

Default value