5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

626

■■■■■■■■

tiamo 2.1 (ProcessLab)

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwanted
EPs.
optimal
Parameter set for all standard titrations; optimized for the most frequent
applications.
fast
For less critical rapid titrations.
User
Editing the individual titration parameters which affect the titration rate.

Measured value acceptance

Note

Will be displayed only if the User option is selected in the Titration
rate
list box.

Measured value drift

Drift for the measured value acceptance during the titration.

Titrando, 855
Range

0.01 ... 99.90 µA/min

Default value

20.00 (slow), 50.00 (optimal), 80.00 (fast) µA/
min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Titrino
Range

0.05 ... 99.90 µA/min

Default value

20.00 (slow), 50.00 (optimal), 80.00 (fast) µA/
min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Min. waiting time

The minimum waiting period is only important for drift-controlled measure-
ments. The measured value is not accepted until the minimum waiting time