beautypg.com

Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 1242

background image

6.3 Configuration data

■■■■■■■■■■■■■■■■■■■■■■

1226

■■■■■■■■

tiamo 2.1 (ProcessLab)

Creating new template or editing template
With or the dialog window New template or Edit tem-
plate
opens for entering a new template or for editing an existing template.

Signal name

Name of the pattern for the output signal.

Input

25 characters

Output signal

Entry of the bit-pattern for the output signal with exactly 14 characters.

It is possible to enter the characters

0 = Line inactive,

1 = Line active,

* = Any line status and

p = Set pulse

Selection

Bit-pattern with exactly 14 characters (0, 1, *,
p)
| **************

Default value

**************

The output lines are numbered from right to left:

Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0

Bit

13 12 11 10 9 8 7 6 5 4 3 2 1 0

Example: ************1* sets the output line 1 to active (= set), which
would result in a Stop command for a connected Titrino, for example.
************0* sets the line to inactive.

Note

We recommend masking the irrelevant output lines with an asterisk * so
as not to modify these line conditions.

Pulse length

Duration of the transmitted pulse.

Range

100 ... 1000 ms

Default value

200 ms

Examples for 765 Dosimat

Parameters

Binary pattern

Function