Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1242
6.3 Configuration data
■■■■■■■■■■■■■■■■■■■■■■
1226
■■■■■■■■
tiamo 2.1 (ProcessLab)
Creating new template or editing template
With
plate opens for entering a new template or for editing an existing template.
Signal name
Name of the pattern for the output signal.
Input
25 characters
Output signal
Entry of the bit-pattern for the output signal with exactly 14 characters.
It is possible to enter the characters
0 = Line inactive,
1 = Line active,
* = Any line status and
p = Set pulse
Selection
Bit-pattern with exactly 14 characters (0, 1, *,
p) | **************
Default value
**************
The output lines are numbered from right to left:
Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Bit
13 12 11 10 9 8 7 6 5 4 3 2 1 0
Example: ************1* sets the output line 1 to active (= set), which
would result in a Stop command for a connected Titrino, for example.
************0* sets the line to inactive.
Note
We recommend masking the irrelevant output lines with an asterisk * so
as not to modify these line conditions.
Pulse length
Duration of the transmitted pulse.
Range
100 ... 1000 ms
Default value
200 ms
Examples for 765 Dosimat
Parameters
Binary pattern
Function