6 meas ipol - fix endpoint evaluation, 5 meas upol, 1 meas upol - overview – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 914: Meas upol
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
898
■■■■■■■■
tiamo 2.1 (ProcessLab)
[New]
Opens the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.10.2, page 859).
[Properties]
Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.10.2, page 859).
[Delete]
Deletes the external measured value selected in the table.
5.6.4.4.6
MEAS Ipol - Fix endpoint evaluation #
Dialog window: Method
▶ MEAS Ipol ▶ Properties... ▶ Additional evaluations ▶ Fix
endpoint evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.
Selection
Measured value | time
Default value
Measured value
Fix value
Value of the fixed endpoint.
Measured value (Titrando, 867, 855)
Range
-2000.0 ... 2000.0 mV
Measured value (Titrino)
Range
-2000 ... 2000 mV
Time
Range
0.0 ... 999999.9 s
5.6.4.5
MEAS Upol
5.6.4.5.1
MEAS Upol - Overview
Dialog window: Method
▶ MEAS Upol ▶ Properties... ▶ MEAS Upol - 'Command
name'
Command for amperometric measurements with selectable polarization
voltage.
Devices
This command can be executed with the following devices: