Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 464
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
448
■■■■■■■■
tiamo 2.1 (ProcessLab)
ascending
Only equivalence points with a positive slope of the titration curve will
be recognized.
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.
off
Equivalence point recognition is switched off.
Evaluation with measured value window (pH)
EP criterion
Criterion for the recognition of equivalence points which is valid for all win-
dows. Equivalence points whose ERC is smaller than the set EP criterion will
not be recognized.
Titrino
Range
0 ... 200
Default value
5
[New]
Opens the dialog window Measured value # in which the parameters for
a new window can be entered (see Chapter 5.6.3.2.2.9, page 454).
[Properties]
Opens the dialog window Measured value # in which the parameters for
the selected window can be edited (see Chapter 5.6.3.2.2.9, page 454).
[Delete]
Deletes the window selected in the table.
Evaluation with volume window (mL)
[New]
Opens the dialog window Volume window # in which the parameters for
a new window can be entered.
[Properties]
Opens the dialog window in which the parameters for the selected window
can be edited (see Chapter 5.6.3.2.2.10, page 455).
[Delete]
Deletes the window selected in the table.