7 set upol - conditioning, Conditioning – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 737
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.1 (ProcessLab)
■■■■■■■■
721
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.4.6.7
SET Upol - Conditioning
Tab: Method
▶ SET Upol ▶ Properties... ▶ Conditioning
Command name
Name of the command.
Input
25 characters
Conditioning
on | off (Default value: off)
Switching conditioning on/off. If conditioning is deactivated all following
parameters are not visible.
Start drift
The determination can only be started when the currently measured drift is
less than the entered start drift entered here. If this is the case, then the
status displayed in the Run subwindow in the Workplace program part
changes to COND READY.
Titrando, 855
Range
1 ... 999 µL/min
Default value
20 µL/min
Drift correction
If this parameter is set to automatic or manual the drift correction ( = Drift
value × Time for drift correction) caused by empty consumption during
titration is deducted from the total consumption. The Time for drift cor-
rection is the period between end of conditioning and end of titration.
Selection
automatic | manual | off
Default value
off
automatic
The drift lastly determined during conditioning will be used as drift value.
manual
The drift value defined under Drift value will be used.