6 load new program version, Load new program version 1 – Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1347
■■■■■■■■■■■■■■■■■■■■■■
7 Devices
tiamo 2.1 (ProcessLab)
■■■■■■■■
1331
Message by E-mail
on | off (Default value: off)
The message is additionally sent to the e-mail address defined under [E-
mail] if this option is activated. The message is sent in text format.
[E-mail]
[E-Mail] opens the Send e-mail window (see Chapter 2.6.1, page 81).
Acoustic signal
on | off (Default value: off)
If this option is enabled, an acoustic signal will be emitted additionally to the
message defined above
Action
The actions can only be edited if monitoring is switched on.
If during monitoring it is found that the validity period has expired then one
of the following actions will be triggered automatically at Start test.
Selection
Record message | Display message | Stop deter-
mination
Default value
Display message
Record message
The message that the validity period has expired will be automatically
saved with the determination.
Display message
A message is displayed and you can choose whether you want to con-
tinue the run or cancel it. If the run is continued then the message that
the validity period has expired will be automatically saved with the deter-
mination.
Stop determination
The running determination will be automatically terminated. The follow-
ing message must be confirmed with [OK].
7.1.6
Load new program version
Dialog window: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device
type' - 'Device name'
▶ General ▶ [Update] ▶ Load new program version
If it is discovered that the device to connect has an old program version not
supported by tiamo, it must be updated. Select the device from the device
table. [Edit]
▶ Properties... opens the properties window. On the Gen-
eral tab, [Update] is now active and can be used to open the Load new
program version dialog window.