5 sample changer - pumping, Pump – Metrohm viva 1.1 (ProLab) User Manual
Page 985
■■■■■■■■■■■■■■■■■■■■■■
8 Manual control
viva 1.1 (for process analysis)
■■■■■■■■
973
External position
Selection of the external position to which the current robotic arm posi-
tion is to be assigned.
Input range
1 - 4
Default value
1
[Assign]
Triggers assignment. During the assignment the cursor is displayed as
hourglass.
8.4.5.5
Sample changer - Pumping
Program part: Manual control
The pumps connected to the tower can be switched on and off here.
Pump 1
Manually control pump 1 or valve 1 on the selected tower.
Operation
Selection whether the pump or valve is to be switched on or off manually
or whether it is to be switched on for a specified duration and then
switched off again automatically.
Selection
On/Off | Duration
Default value
On/Off
Duration
Entry of the duration during which the pump/valve should remain
switched on. This field is only displayed for Operation = Duration.
Input range
1 - 999,999 s
Default value
8 s
[Start]
Starts pump 1. The running time since start is shown in the status display.
[Stop]
Stops pump 1.
Pump 2
Manually control pump 2 or valve 2 on the selected tower.
Operation
Selection whether the pump or valve is to be switched on or off manually
or whether it is to be switched on for a specified duration and then
switched off again automatically.