6.3 Configuration data
■■■■■■■■■■■■■■■■■■■■■■
760
■■■■■■■■
viva 1.1 (for process analysis)
6.3.3
Templates
6.3.3.1
Templates - Input lines
6.3.3.1.1
Managing templates for input lines
Dialog window: Configuration
▶ Tools ▶ Templates ▶ Input lines... ▶ Templates
for input lines
Client-specific bit patterns for scanning remote input signals can be
defined in the Templates for input lines dialog window and selected
with the SCAN command and in manual control. The table with the
defined templates cannot be edited, but it can be sorted according to the
selected column in ascending or descending order by clicking on the col-
umn title.
The [Edit] menu below the list of templates or a right-click on the table
opens a menu with the following menu items:
New...
Create a new template (see Chapter 6.3.3.1.2, page 760).
Properties...
Edit the selected template (see Chapter 6.3.3.1.2, page 760).
Delete
Deletes the selected template.
6.3.3.1.2
Editing templates for input lines
Dialog window: Configuration
▶ Tools ▶ Templates ▶ Input lines... ▶ Templates
for input lines
▶ New... / Properties... ▶ New template / Edit template
Signal name
Name of the template for the input signal.
Entry
25 characters
Input signal
Entry of the bit pattern for the input signal with exactly 8 characters. It is
possible to enter the characters
0 = Line inactive,
1 = Line active and
* = Any line status.
Selection
Bit pattern with 8 characters (0, 1, *) | ********
Default value
********
The input lines and bits are numbered from right to left:
Input 7 6 5 4 3 2 1 0
Bit
7 6 5 4 3 2 1 0