2 meas ref - general/hardware – Metrohm viva 1.1 (ProLab) User Manual
Page 523
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
511
Parameters
The parameters for the MEAS Ref command are set on the following
tabs:
■
General/Hardware
Parameters for instrument.
■
Measuring parameters
Parameters for setting the measurement procedure.
Command variables
The following command variables are generated by the MEAS Ref com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':
Identifi-
cation
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
never been started
DBL
Total duration for the processing of the command in s
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
not valid (variable not available) = Command has
not yet been started
STY
Type of stop with which the command was stopped: 1 =
finished normally, 0 = stopped manually with [Quit]
or [Stop]
5.5.2.7.2.2
MEAS Ref - General/Hardware
Tab: Method
▶ MEAS Ref ▶ Properties... ▶ MEAS Ref - 'Command name' ▶ Gen-
eral/Hardware
Command name
Name of the command.
Entry
25 characters
The general parameters for the instrument are defined on this tab.
Device
Device name
Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.