5 cp - post-treatment, Post-treatment – Metrohm viva 1.1 (ProLab) User Manual
Page 660
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
648
■■■■■■■■
viva 1.1 (for process analysis)
Measuring interval
Measuring interval for data recording during plating or during the OCP
measurement (see Glossary, page 1055).
Input range
0.1 - 60.0 s (Increment: 0.1)
Default value
1.0 s
Number of measuring points
Shows the number of measuring points.
NOTE
The actual potential measurement takes place during the last 100 ms of
the measuring interval.
5.5.2.11.4.5
CP - Post-treatment
Tab: Method
▶ CP ▶ Properties... ▶ Post-treatment
Command name
Name of the command.
Entry
25 characters
Cleaning
Cleaning potential
Potential that is applied to the electrodes during the cleaning time. This
potential can be used e.g. for the electrochemical cleaning of solid-state
electrode surfaces that have been contaminated with reaction products of
electrode redox processes.
Input range
–5.000 - 5.000 V (Increment: 0.001)
Selection
off
Default value
off
Cleaning time
Waiting time during which the cleaning potential is applied to the elec-
trodes. The field can only be edited for Cleaning potential ≠ off.
Input range
0.0 - 9,999.9 s (Increment: 0.1)
Default value
0.0 s