3 Workplace
viva 1.1 (for process analysis)
■■■■■■■■
137
■
Instrument (module) and rack check
Check whether the instruments (or instrument types), instrument mod-
ules (e.g. tower, etc.) and racks requested in the instrument-specific
commands are present and available or assigned. If an instrument has
not yet been assigned in a command, then the Device assignment
dialog window will open for the assignment of the instrument.
■
Checking solutions
Check whether the solutions requested in the instrument-specific com-
mands are present and available.
■
Checking dosing units
Check whether the dosing units requested in the instrument-specific
commands are present and available.
■
Checking sensors
Check whether the sensors requested in the instrument-specific com-
mands are present and available.
NOTE
In the status READY, the start test can also be triggered manually with
the symbol or the Tools
▶ Run test menu item. In addition, the
sample data test will also be carried out afterwards.
4 - Method start
The following checks and actions are triggered with the method start:
■
Checking sample data
Check whether the sample data entered corresponds to the defined
sample data variables and whether it is valid.
■
Assigning determination ID
A unique determination ID is assigned for the determination.
■
Reserving instruments (or instrument modules)
The instruments or their function units used by the instrument-depen-
dent commands of the method are reserved for the duration of the
determination, i.e. they can neither be used on other workplaces nor
operated manually.
■
Increasing sample number
The Sample number is increased by +1.
5 - Main run
The following actions can be carried out during processing of the loaded
method: